A system for automatically calibrating oscilloscopes using MATLAB software
1
School of Mechanical Engineering, Shijiazhuang Tiedao University, Shijiazhuang City, Hebei Province 050043, China
*
Author to whom correspondence should be addressed.
SIA 1(1):11; https://doi.org/10.12410/sia0101011
Received: 13 March 2024 / Accepted: 20 March 2024 / Published Online: 22 April 2024
Abstract
This article proposes a system for automatic calibration of general-purpose digital oscilloscopes using MATLAB software. The computer system is connected to the oscilloscope calibrator 9500B through a GPIB interface card, and the digital oscilloscope interface is connected through a USB cable. The control program is written using SCPI instructions, and ten functions are measured and verified item by item. After the test is completed, the test data is output to complete the automatic verification process. This system is optimized to address the limitations of existing oscilloscope calibration systems on the market, which helps to improve the efficiency and accuracy of oscilloscope calibration.
Copyright: © 2024 by Jin, Si, Shen, Wu and Chen. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY) (Creative Commons Attribution 4.0 International License). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
Share and Cite
ACS Style
Jin, P.; Si, L.; Shen, Y.; Wu, H.; Chen, J. A system for automatically calibrating oscilloscopes using MATLAB software. Scientific Innovation in Asia, 2023, 1, 11. doi:10.12410/sia0101011
AMA Style
Jin P, Si L, Shen Y et al.. A system for automatically calibrating oscilloscopes using MATLAB software. Scientific Innovation in Asia; 2023, 1(1):11. doi:10.12410/sia0101011
Chicago/Turabian Style
Jin, Peng; Si, Longhuan; Shen, Yaliang; Wu, Hangbing; Chen, Junjie 2023. "A system for automatically calibrating oscilloscopes using MATLAB software" Scientific Innovation in Asia 1, no.1:11. doi:10.12410/sia0101011
Article Metrics
Article Access Statistics
References
- [1] Fumin Lu, Zhiqiang Zhan (2004). Exploration of digital oscilloscope scanning time calibration method. Journal of Electronic Measurement and Instrumentation, Supplement, 527-530.
- [2] Fluke Corporation (2002). Fluke Corporation User's handbook for model 9500B high performance calibrator operation and performance,100-130.
- [3] Tektronix Ltd.(2004) DPO4000 series digital oscilloscopes user manual, 78-143.
- [4] Fan Qin ,Gao Wei. (2011 ) Instrumental programmable measurement based on VISA library and SCPI command. Modern Electronic Technology, 34,118-120.
- [5] Weimin Guo, Xiaoli Deng, Li Li. (2004) Research on Verification Methods for Digital Oscilloscopes. Information and Electronic Engineering, 2, 65-69.
- [6] Ben Liang, Xungao Li, Jiang Lin. (2010) Signal Processing Analysis and Research Based on MATLAB and LabVIEW. Experimental Science and Technology, 8, 39-41.